PROGRAMMER'S GUIDE Page 97
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
CyberAmp 380 Command Set Details
In the descriptions that follow, it is assumed that the <Carriage Return> character is sent at the end of
the line, and for clarity, it will not be included in the command string.
In all command strings, spaces and leading zeros are ignored.
"C" Input Coupling
Format
ATdCnpfff
where
d = device # (0-9)
n = channel number (1-8) or "?"
p = polarity to set, positive (+) or negative (-)
fff = frequency of the input coupling
Examples
AT2C1+100
Set device 2, channel #1, positive input = 100Hz
AT2C1-300
Set device 2, channel #1, negative input = 300Hz
AT2C1-GND
Set device 2, channel #1, negative input = GND
AT2C1+DC
Set device 2, channel #1, positive input = DC coupled
AT3C?
Return all possible input coupling values of device 3
Response:
GND DC 0.1 1 10 30 100 300
Notes
The data field can have 1-3 characters, and uses which may have leading zero suppression (i.e., 001 is
the same as 1).
The data field is the coupling frequency. The coupling can be one of six possible AC coupling
frequencies, DC coupling or grounded input (GND). The list of possible values is:
300, 100, 30, 10, 1, 0.1 Hz, DC, and GND
If "?" is supplied after the "C", then all possible coupling values are returned, separated by one or
more spaces.