Pagina 1 - MUSCLE STRIP MYOGRAPH SYSTEM

11User ManualVersion 1.2MUSCLE STRIP MYOGRAPH SYSTEMMODEL 820MS

Pagina 2

10 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALUNPACKING THE MYOGRAPH SYSTEMTake a few minutes to carefully inspect your new Muscle Strip My

Pagina 3 - TRADEMARKS

11CHAPTER 1 - SYSTEM OVERVIEW1.1 Myo-Interface front panelCHAPTER 11.2 Interface Rear PanelPower indicatorMyo-Interface displayValve buttonsON/OFF swi

Pagina 4 - CONTENTS

12 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL1.3 Muscle Strip myograph unitMicrometerConnection to Myo-InterfaceMyograph clamp support con

Pagina 5 - INTRODUCTION

13CHAPTER 2 - SETTING-UP2.1 The complete Muscle Strip Myograph System - 820MS2.2 Setting up step-by-stepThis chapter contains a complete step-by-step

Pagina 6

14 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALFigure 2.2 - Oxygen supply and suction connection Figure 2.3 - Chamber cover Figure 2.4 - C

Pagina 7 - APPROVALS

15CHAPTER 3CHAPTER 3 - THE INTERFACE MENUSMenus on the 820MS interface are all accessible by a touch screen. To ac-cess a menu, simply touch the scre

Pagina 8 - CERTIFICATE OF CONFORMITY

16 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL3.2 Interface Menus – Calibrated ModeIn calibrated mode, the internal amplier is engaged and

Pagina 9 - ABOUT THIS MANUAL

17CHAPTER 3Zero MenuThis menu is used to zero the output of the transducers. When using a data acquisition program like LabChart by AD Instruments ®,

Pagina 10 - UNPACKING THE MYOGRAPH SYSTEM

18 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL1. Force calibration NOTE: EVERYTIME A FORCE CALIBRATION IS PERFORMED THE MEAS-UREMENT RANGE

Pagina 11 - CHAPTER 1 - SYSTEM OVERVIEW

19CHAPTER 3Step 3 Initiates the heating process for the chambers. In order to perform an accurate calibration the transducers must be heated to the d

Pagina 12

2 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL

Pagina 13 - CHAPTER 2 - SETTING-UP

20 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL2. Valve delay:Pressing “VALVE DELAY” in the SETTINGS menu will allow the user to modify the

Pagina 14 - (optional item)

21CHAPTER 34. Measurement Range:The MEASUREMENT RANGE sub-menu in SETTINGS determines the maxi-mum force capacity of the transducer. The factory set

Pagina 15 - 15CHAPTER 3

22 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALII. RECORD OUTPUT SELECT The RECORD OUTPUT SELECT option allows the end-user to select the

Pagina 16 - Multi Myograph System

23After a few seconds, the DMT 820MS Direct Force Output display will ap-pear. When in Direct Mode, the ACTUAL FORCE display will not be shown, as is

Pagina 17 - SETTINGS

24 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALCHAPTER 4 - THE MUSCLE STRIP MYOGRAPH UNITChapter 4 contains a complete explanation of how to

Pagina 18 - Force calibration chamber 4:

25 Figure 4.1 - 4.2 Weight calibration kit shown in place on a single myograph chamberThe 2 g weight is placed on this panThe tip of the transducer

Pagina 19 - 19CHAPTER 3

26 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL4.3 Force Transducer ReplacementIf the force transducer breaks and needs to be replaced, foll

Pagina 20 - Chamber 4: 5 Sec

274. The replacement transducer will be shipped with the new transducer inside a new transducer house.5. Place a VERY small amount of vacuum grease

Pagina 21 - INTERFACE SETTINGS

28 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL4.4 Myograph MaintenanceThe Muscle Strip Myograph System - 820MS is a very delicate and sophi

Pagina 22

294.5 Cleaning the myographDMT strongly recommends that the myograph chambers and surrounding areas are cleaned after each experiment.At the end of ea

Pagina 23 - Pres enter to zero

3MUSCLE STRIP MYOGRAPH SYSTEMMODEL 820MSTRADEMARKSPowerLab® and LabChart® are registered trademarks of ADInstruments Pty Ltd. The names of specic rec

Pagina 24

30 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALCHAPTER 5 - THE MOUNTING SUPPORTS AND MOUNTING IN THE 820MS CHAM-BEREach chamber contains sup

Pagina 25 - CHAPTER 4

31APPENDIX 1APPENDIX 1 - SYSTEM SPECIFICATIONS Technical specicationsVessel size: >3 mm - 19 mmChamber: Four individual chambers Chamber mat

Pagina 26

3232820MS/02/2013Aalborg Hospital South • Academic Medical Center Amsterdam • Academy of Sciences of the Czech Republic • Actelion Pharmaceuticals Ltd

Pagina 27

4 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALCONTENTSTrademarks ...

Pagina 28 - 4.4 Myograph Maintenance

5INTRODUCTIONThe Muscle Strip Myograph System - 820MS represents the rst state-of-the-art 4-channel myograph system for muscle strips of up to 15 mm

Pagina 29 - 4.5 Cleaning the myograph

6 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALSAFETYThe Muscle Strip Myograph System - 820MS has been designed for use in teaching and resea

Pagina 30

7EMC / EMIThis equipment has been tested and found to comply with the limits for a Class B Digital device, pursuant to part 15 of the FCC rules. These

Pagina 31 - Optional accessories

8 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALCERTIFICATE OF CONFORMITYDMT A/S, Skejbyparken 152, 8200 Aarhus N., Denmark, hereby declares i

Pagina 32 - 820MS/02/2013

9ABOUT THIS MANUALThis manual contains a complete list of procedures that describe how to install, maintain and use the Muscle Strip Myograph System –