11User ManualVersion 1.2MUSCLE STRIP MYOGRAPH SYSTEMMODEL 820MS
10 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALUNPACKING THE MYOGRAPH SYSTEMTake a few minutes to carefully inspect your new Muscle Strip My
11CHAPTER 1 - SYSTEM OVERVIEW1.1 Myo-Interface front panelCHAPTER 11.2 Interface Rear PanelPower indicatorMyo-Interface displayValve buttonsON/OFF swi
12 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL1.3 Muscle Strip myograph unitMicrometerConnection to Myo-InterfaceMyograph clamp support con
13CHAPTER 2 - SETTING-UP2.1 The complete Muscle Strip Myograph System - 820MS2.2 Setting up step-by-stepThis chapter contains a complete step-by-step
14 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALFigure 2.2 - Oxygen supply and suction connection Figure 2.3 - Chamber cover Figure 2.4 - C
15CHAPTER 3CHAPTER 3 - THE INTERFACE MENUSMenus on the 820MS interface are all accessible by a touch screen. To ac-cess a menu, simply touch the scre
16 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL3.2 Interface Menus – Calibrated ModeIn calibrated mode, the internal amplier is engaged and
17CHAPTER 3Zero MenuThis menu is used to zero the output of the transducers. When using a data acquisition program like LabChart by AD Instruments ®,
18 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL1. Force calibration NOTE: EVERYTIME A FORCE CALIBRATION IS PERFORMED THE MEAS-UREMENT RANGE
19CHAPTER 3Step 3 Initiates the heating process for the chambers. In order to perform an accurate calibration the transducers must be heated to the d
2 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL
20 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL2. Valve delay:Pressing “VALVE DELAY” in the SETTINGS menu will allow the user to modify the
21CHAPTER 34. Measurement Range:The MEASUREMENT RANGE sub-menu in SETTINGS determines the maxi-mum force capacity of the transducer. The factory set
22 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALII. RECORD OUTPUT SELECT The RECORD OUTPUT SELECT option allows the end-user to select the
23After a few seconds, the DMT 820MS Direct Force Output display will ap-pear. When in Direct Mode, the ACTUAL FORCE display will not be shown, as is
24 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALCHAPTER 4 - THE MUSCLE STRIP MYOGRAPH UNITChapter 4 contains a complete explanation of how to
25 Figure 4.1 - 4.2 Weight calibration kit shown in place on a single myograph chamberThe 2 g weight is placed on this panThe tip of the transducer
26 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL4.3 Force Transducer ReplacementIf the force transducer breaks and needs to be replaced, foll
274. The replacement transducer will be shipped with the new transducer inside a new transducer house.5. Place a VERY small amount of vacuum grease
28 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUAL4.4 Myograph MaintenanceThe Muscle Strip Myograph System - 820MS is a very delicate and sophi
294.5 Cleaning the myographDMT strongly recommends that the myograph chambers and surrounding areas are cleaned after each experiment.At the end of ea
3MUSCLE STRIP MYOGRAPH SYSTEMMODEL 820MSTRADEMARKSPowerLab® and LabChart® are registered trademarks of ADInstruments Pty Ltd. The names of specic rec
30 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALCHAPTER 5 - THE MOUNTING SUPPORTS AND MOUNTING IN THE 820MS CHAM-BEREach chamber contains sup
31APPENDIX 1APPENDIX 1 - SYSTEM SPECIFICATIONS Technical specicationsVessel size: >3 mm - 19 mmChamber: Four individual chambers Chamber mat
3232820MS/02/2013Aalborg Hospital South • Academic Medical Center Amsterdam • Academy of Sciences of the Czech Republic • Actelion Pharmaceuticals Ltd
4 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALCONTENTSTrademarks ...
5INTRODUCTIONThe Muscle Strip Myograph System - 820MS represents the rst state-of-the-art 4-channel myograph system for muscle strips of up to 15 mm
6 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALSAFETYThe Muscle Strip Myograph System - 820MS has been designed for use in teaching and resea
7EMC / EMIThis equipment has been tested and found to comply with the limits for a Class B Digital device, pursuant to part 15 of the FCC rules. These
8 MUSCLE STRIP MYOGRAPH SYSTEM - MODEL 820MS USER MANUALCERTIFICATE OF CONFORMITYDMT A/S, Skejbyparken 152, 8200 Aarhus N., Denmark, hereby declares i
9ABOUT THIS MANUALThis manual contains a complete list of procedures that describe how to install, maintain and use the Muscle Strip Myograph System –